我科自 1997年引进日本日立公司生产的HITACHI7170A全自动生化分析仪 ,经二年多的使用碰到一些问题 ,在这里一并提出 ,以供同道们参考。1 仪器内通信异常 一个时期以来 ,我们经常碰到仪器提示 :与操作部通信时检测出同步通信 (定时通信 )异常 ,仪器自动停止重新启动?
In 1997, our department introduced the HITACHI 7170A automatic biochemical analyzer manufactured by Hitachi, Japan. After more than two years of use, we encountered some problems that were raised here for reference by our fellow practitioners. 1 Intra-equipment communication is abnormal For a period of time, we frequently encounter instrument prompts: When communicating with the operation unit, abnormal communication (timed communication) is detected, and the instrument automatically stops and restarts.