A convenient derivatization method of amino acids with l-fluoro-2,4-dimtrobenzene as reaction reagent and a separation system were described. The derivative amino acids were separated on a specific chemically bonded phase column with a simple linear gradient elution consisting of aqueous buffer and methanol. The eluate was detected by common ultraviolet absorption detector at 360 nm. The detection limits of amino acids were as low as 10 picomole. This method has been successfully applied to assay amino acid injection liquor used in hospital. It has good repro-ducibility and precision. The procedures avoid the requirements of particular derivative equipment and analyzer employed in conventional amino acid analysis.
A convenient derivatization method of amino acids with l-fluoro-2,4-dimtrobenzene as reaction reagent and a separation system were described. The derivative amino acids were separated on a specific chemically bonded phase column with a simple linear gradient elution consisting of aqueous buffers The assay has been detected by common ultraviolet absorption detector at 360 nm. The detection limits of amino acids were as low as 10 picomole. This method has been successfully applied to assay amino acid injection liquor used in hospital. It has good repro- ducibility and precision. The procedures avoid the requirements of particular derivative equipment and analyzer employed in conventional amino acid analysis.