当硒的含量小于0.01ppm时,用荧光分光光度法测定硒,520nm处的检测峰将受到干扰.用二阶导教对光谱进行处理,可消除这些干扰.该方法的检测限达0.1ppb,回收率为98.8%-101%,变异系数为0.057.
When the content of selenium is less than 0.01 ppm, the detection peak at 520 nm will be disturbed when selenium is determined by fluorescence spectrophotometry. Using second-order teaching to process the spectrum eliminates these disturbances. The detection limit of the method is 0.1ppb, the recovery rate is 98.8% -101%, and the coefficient of variation is 0.057.