在 X 线 CT 机中较易出现故障的元件之一是 X 线高压发生电路中的调压器。岛津 SCT-100N-2S 型头部 CT 机中的调压器原理图如图所示。该调压器工作频繁,CT 机每扫描一个断层就工作一次(即调压器输出轮由电机带动由调压器的低端自动调到高端,当 X 光管球曝光结束后又自动退回到低端)。而且都是满负荷运行,平均每天都要工作几百次以上,所以它的故障率较高。几年来我院的岛津 SCT-100N-2S型 CT 机的调压器多次出现故障,以下就是具体
One of the more troublesome components in an X-ray CT machine is a voltage regulator in the X-ray high voltage generating circuit. Shimadzu SCT-100N-2S-head CT machine schematic diagram shown in the figure. The regulator work frequently, CT scan a fault on the work once (that is, the output of the regulator by the motor driven by the low end of the regulator automatically transferred to the high end, when the X-ray tube after the end of exposure is automatically returned to Low end). And they are all running at full capacity, working on average hundreds of times a day or more, so it has a higher failure rate. In recent years, our hospital Shimadzu SCT-100N-2S-type CT pressure regulator repeatedly failure, the following is specific