论文部分内容阅读
WZZ—2型自动旋光仪在使用中,有时出现尾数两位数字跳字跳动,有时很容易烧Na灯泡,我院曾在一个月内烧坏6只Na灯泡,该厂没有给电路图,因此,我根据印刷电路图(如图1),这是一个串联稳压电源.(虚线连接的部分是改动添加部分)即没有取样,也没有保护电路.T
In the use of the WZZ-2 automatic polarimeter, there are occasional two-digit digital jumps in the mantissa. It is sometimes very easy to burn Na bulbs. Our hospital burned six Na bulbs within a month. The factory did not give circuit diagrams. Therefore, I according to the printed circuit diagram (Figure 1), this is a series regulated power supply. (The part connected by the dotted line is the modification addition part) ie no sampling and no protection circuit. T