第八章非均匀磁场仪器以上介绍的单聚焦式和双聚焦式静态质谱仪器都采用均匀磁场,一般只考虑径向聚焦情况。这些仪器的设计理论较为完善,磁铁结构比较简单,性能上也常能满足实验要求,目前应用较广。然而,即使磁铁极靴加工得十分平整、安装得十分对称,也只能保证中心部份具有均匀磁场,而边缘部份则是表现为非均匀磁场的弥散场(图8—1)。弥散磁场具有径向分量和轴向分量,如果处理得当,可以起着径向和轴
Chapter 8 Inhomogeneous Magnetic Field Instruments The single-focus and dual-focus static mass spectrometers described above all use a uniform magnetic field. Generally only radial focusing is considered. The design theory of these instruments is more perfect, the magnet structure is relatively simple, the performance can often meet the experimental requirements, the current wider application. However, even if the pole pieces are machined to be very flat and perfectly symmetrical, only the central part will have a uniform magnetic field while the edge parts will appear as a diffuse magnetic field with a non-uniform magnetic field (Figure 8-1). Diffusive magnetic fields have both radial and axial components and, if handled properly, can act as radial and axial