故障现象一:一开机,显示屏上所有字符与波形稳定不住,向右跑动。故障分析与处理:从屏幕现象可看出,这可能是行同步频率超出同步范围所致。打开机盖,仔细观察主板上有一可调节
【机 构】
:
马鞍山市人民医院器械科!安徽马鞍山243000
论文部分内容阅读
故障现象一:一开机,显示屏上所有字符与波形稳定不住,向右跑动。故障分析与处理:从屏幕现象可看出,这可能是行同步频率超出同步范围所致。打开机盖,仔细观察主板上有一可调节电位器AIR21,此电位器即行同步频率范围调节电位器,可用来调节行同步范围,用一细长一字螺丝刀小心调
Symptom 1: When the camera is turned on, all the characters and waveforms on the screen cannot be stabilized and run to the right. Fault analysis and processing: As can be seen from the screen phenomenon, this may be caused by the line synchronization frequency exceeding the synchronization range. Open the cover and carefully observe that there is an adjustable potentiometer AIR21 on the mainboard. This potentiometer is the line synchronization frequency range adjustment potentiometer, which can be used to adjust the line synchronization range. Carefully adjust with a slender flat-head screwdriver.
其他文献
曾几何时,我们开始热衷“模仿”,模仿歌星、影星、笑星……娱乐节目中铺天盖地的“模仿秀”也不断入侵我们的视听。现在,有这样一群人,“模仿”成为他们的一种生存状态:他们
请下载后查看,本文暂不支持在线获取查看简介。
Please download to view, this article does not support online access to view profile.
1 故障现象开机液晶显示屏显示“STOPBLOODPUMP”,并且屏幕显示连续冲洗计时时间,听声音机器确在冲洗程序状态下运行,打开冲洗开关,能正常冲洗。2 故障分析(1)冲洗开关接点不正常,引起程
山西籍名作家兼学者韩石山先生,在《我不配上中国人民大学的教科书》(《文学自由谈》2003年第3期)一文中,半调侃半义愤地反驳了据说是中国人民大学某教科书对他的批评,足见韩
故障现象 :无论是用 3D吊管摄片或Sireskop3胃肠床点片 ,按正常曝光条件选择kV、mAs照片显影淡薄 ,无法用于临床诊断。故障分析与检查 :根据故障现象分析 ,床上管和吊管在摄片同时出现影像