KB-400型400毫安X线机为全波整流、双球管、双焦点的诊断X线机。该机控制电路大都采用2H型、3H型及DZ-644型等不同类型的继电器。由于使用维护不当或因制造质量不良等原因,继电器发生故障是常见的,严重的影响了X线检查工作。为此,我们将该机中的TDR和RA继电器进行了改装处理,获得良好效果。现将改进方法介绍如下,供参考。一、延时继电器该机延时继电器TDR是在通电后,电动机开始运转,经0.8~1.2秒后接点闭合,
The KB-400 400 mA X-ray machine is a full-wave rectification, dual-tube, dual-focus diagnostic X-ray machine. Most of the control circuits of this machine adopt different types of relays such as 2H type, 3H type and DZ-644 type. Due to improper use and maintenance or due to poor manufacturing quality, failure of relays is common and seriously affects X-ray inspection work. For this purpose, we modified the TDR and RA relays in this machine and achieved good results. The improved method is described below for reference. First, the delay relay The machine delay relay TDR is after the power, the motor starts running, after 0.8 to 1.2 seconds after the contact closure,