论文部分内容阅读
We studied optimal conditions for the determination of oxygen consumption rate (OCR).The assay, with a working range up to 35 mmol/L, had an intra-assay coefficient of variation of 0. 7%~3. 52%, and the inter-assay coefficient of variation was 4. 86%. In spiking experiments, the recovery range was 98. 9%~101 %. The assay was compared with CX3 (Beckman Co.). A linear regression analysis yielded a slope of 1. 05, an inercept of 3. 82 and a correlation coefficient of 0. 999. The serum of turbidity above + + had a negative interference on the assay (>3. 4%). The obvious interference of hemolysis, icterus and vitamin C to the assay were not observed.
We studied optimal conditions for the determination of oxygen consumption rate (OCR). The assay, with a working range up to 35 mmol / L, had an intra-assay coefficient of variation of 0.7-3.22%, and the The assay was compared with CX3 (Beckman Co.). A linear regression analysis yielded a slope of 1. The inter-assay coefficient of variation was 4.86%. In spiking experiments, the recovery range was 98.9% ~ 101%. 05, an inercept of 3. 82 and a correlation coefficient of 0. 999. The serum of turbidity above + + had a negative interference on the assay (> 3.4%). The obvious interference of hemolysis, icterus and vitamin C to the assay were not observed