We have tested and analyzed the properties of two-dimensional Position-Sensitive-silicon-Detector (PSD) with new integrated preamplifiers. The test demonstrates that the best position resolution for 5.5 MeV α particles is 1.7 mm (FWHM), and the best energy resolution is 2.1%, which are notably better than the previously reported results. A scaling formula is introduced to make the absolute position calibration.
We have tested and analyzed the properties of two-dimensional Position-Sensitive-silicon-Detector (PSD) with new integrated preamplifiers. The test demonstrates that the best position resolution for 5.5 MeV alpha particles is 1.7 mm (FWHM), and the best energy resolution is 2.1%, which are notably better than the previously reported results. A scaling formula is introduced to make the absolute position calibration.