论文部分内容阅读
目的了解三门峡市医用诊断X射线设备的的影像质量控制现状,为进一步开展质量保证工作打下基础。方法使用X射线机质控检测装置对三门峡市181台X射线机进行了质控检测。结果各项指标均合格的X射线摄影和透视设备分别占同类检测总台数的47.7%和48.6%。结论三门峡市医用诊断X射机影像质量控制现状仍存在很多问题,定期进行影像质量控制检测是十分必要的。
Objective To understand the status quo of image quality control of medical diagnostic X-ray equipment in Sanmenxia and lay the foundation for further quality assurance. Methods Quality control tests were carried out on 181 X-ray machines in Sanmenxia City using X-ray quality control testing equipment. Results The qualified X-ray radiography and fluoroscopy equipment accounted for 47.7% and 48.6% of the total number of similar tests. Conclusion There are still many problems in the quality control of medical diagnostic X-ray machine in Sanmenxia City. It is very necessary to conduct regular inspection of image quality control.