现将TALYSURF 5表面光洁度测量仪常见故障及其排除分述如下: (1)电源部分 Talysuf 5系统中,故障最多的就是电源部分,而功率放大器的故障又居首位。若面板上某些指示灯不亮或全部指示灯都不亮,或移笔器失灵,不能使pick-up position的绿色光带左右移动多半是电源部分出了故障。可以先检查后面板上的保险管(当心触电!最好先拔掉电源插头)。如果都是完好的,则可检查机壳内的保险管。打开仪器左面盖板(近电源开关侧)仪器后部有一个电源变压器,变压器后面有一块小板,即为保险管板(此处距离电源电压选择器
Now TALYSURF 5 surface finish meter common faults and their exclusions are described below: (1) Power part of the Talysuf 5 system, the fault is the power part of the most, and the power amplifier failure first place. If some of the indicators on the panel are not lit or all the lights are off, or the printer fails, you can not make the green light of the pick-up position move left and right most of the time, the power part of the failure. You can first check the fuse on the back panel (beware of electric shock! It is best to unplug the power plug). If all are intact, you can check the fuse inside the case. Open the instrument on the left cover (near the power switch side) there is a power transformer in the rear of the instrument, there is a small plate behind the transformer, which is the fuse board (here from the power supply voltage selector