论文部分内容阅读
Purpose: To inspect and compare the image quality of an aspheric intraocular lens (IQ, Alcon) with those of conventional monofocal silicone and acrylic intraocular lens and multifocal intraocular lens (Array).Methods: The IOLs were tested in the eye model, which was designed to be optically equivalent to the theoretical eye model. The eye model is a combination of a spherical photographic lens with 35 mm focal length (IOL put in a water cell) and a charge coupled device (CCD) camera. The images constructed by the lenses are observed on a monitor of personal computer and the contrasts of the images are analyzed by using commercial image processing software. SHARP value is used to measure and estimate image definition.Results: The images constructed by changing the diameter of aperture stop and IOL. Observed by this eye model, the image definition of aspheric intraocular lens (IQ, Alcon) is better than others.Discussion: The proposed eye model is useful for testing functional vision and for inspecting the differences of intraocular lens.