同其它放射诊断设施一样,已开始在乳腺X线照相术的领域内,借助物理技术参数使照片质量标准化。并有规律地定期检测。乳腺X线照相术的质量检查分为三类:1.验收检测;2.现状检测;3.稳定性检测。验收检测应在乳腺X线照相装置安装或大修后施行。验收检测的间隔期限按1988年1月生效的放射法规第一段的要求执行。由于验收检测必须具备熟悉仪器每一部件的技术结构和功能的知识,因此只有在制造厂家的合作下才能进行。与之相反,稳定性检测则较简便,是测定与照片质量和射线曝光有关的重要参数。放射法规要求稳定性检测必须“在有规律的周期内进行,最少为一月一次”(16条第2款)。
Like other radiological diagnostic facilities, the quality of photos has been standardized with the aid of physical and technical parameters in the field of mammography. And regular regular testing. The quality of mammography is divided into three categories: 1. Acceptance testing; 2. Status quo; 3. Stability testing. Acceptance testing should be performed after mammography installation or overhaul. The intervals between acceptance tests are to comply with the requirements of the first paragraph of the Radio Regulations, which came into force in January 1988. Since acceptance testing must be familiar with the technical structure and function of each part of the instrument, it can only be done in cooperation with the manufacturer. In contrast, stability testing is simpler and is an important parameter for determining the quality of a photo and exposure to radiation. Radiological regulations require stability testing to be “at regular intervals, at least once a month” (Article 16, paragraph 2).