论文部分内容阅读
目的探讨生化分析仪的两个模块检测大鼠血清的差异。方法在同一实验室相同的运行环境下,使用日立7600生化分析仪的D和P两个不同模块,同时检测大鼠血清尿酸(UA)、总胆固醇(TC)、甘油三酯(TG)和高密度脂蛋白胆固醇(HDL-C)。结果日立7600生化分析仪D和P两个模块检测大鼠血清结果差异无统计学意义(尿酸t=1.751,总胆固醇t=-1.318,甘油三酯t=-1.472,高密度脂蛋白胆固醇t=1.941,P>0.05),两个模块测定大鼠血清UA、TC、TG和HDL-C相关系数r分别是0.993、0.992、0.999、0.996。结论日立7600生化分析仪的两个不同模块检测大鼠血清UA、TC、TG和HDL-C结果有良好的一致性。
Objective To investigate the differences between the two modules of biochemical analyzer in detecting rat serum. Methods Two different D and P modules of Hitachi 7600 biochemical analyzer were used in the same operating environment of the same laboratory to detect serum uric acid (UA), total cholesterol (TC), triglyceride (TG) and high Density lipoprotein cholesterol (HDL-C). Results Hitachi 7600 biochemical analyzer D and P two modules to detect rat serum results were no significant difference (uric acid t = 1.751, total cholesterol t = -1.318, triglyceride t = -1.472, high-density lipoprotein cholesterol t = 1.941, P> 0.05). The correlation coefficients of serum UA, TC, TG and HDL-C in the two models were 0.993, 0.992, 0.999 and 0.996, respectively. Conclusion The two different modules of Hitachi 7600 Biochemical Analyzer have good consistency in detecting serum UA, TC, TG and HDL-C in rats.