一台瑞典金宝GAMBROAK100透析机开机自检不通过,提示需维修的扳手指示灯亮,同时提示多达十三个错误代码。根据错误代码查询,提示有多个泵和电磁阀失效,同时还提示A/D变换锁定于“忙”状态和CPU软件错误。先依照提示检查各泵和阀,未见异常;FM板上...
A Swedish Campbell GAMBROAK100 dialysis machine did not pass the self-test, prompting the need to repair the wrench indicator light, and prompts up to thirteen error codes. According to the error code query, it indicates that there are multiple pumps and solenoid valves failing. It also indicates that the A/D converter is locked in the “busy” state and the CPU software error. First check the pumps and valves according to the prompts. No abnormality is found. On the FM board...