北京F78—ⅢA型X线机原点片装置在进行食道造影检查时,只能用254mm×305mm(10″×12″)的胶片进行二分格点片。由于食道钡餐检查需适时点片,而原装置一次仅能点片两张,若要继续点片得更换一次暗匣,这往往容易失去抢拍的最佳时机。虽然原装置设有四次曝光档,但由于曝光窗较小,不宜于食道检查。为此,我们重新设计了一块附加装置,与该型机器原遮光板配合,使曝光窗变成73mm×278mm,即可在280mm×356mm(11″×14″)胶片上连续四分格点片。制作方法如下。
Beijing F78-IIIA X-ray machine original spot device can only use 254mm × 305mm (10′′ × 12′′) film for two-dimension lattice inspection during the esophageal radiography examination. Because esophageal barium meal examination needs to be spotted in time, and the original device can only spot two at a time, if you want to continue to change the spot, you may lose the best time to snap. Although the original device has four exposures, it is not suitable for esophageal examination due to the small exposure window. For this reason, we redesigned an additional device to cooperate with the original shading plate of this type of machine so that the exposure window would be 73mm x 278mm, which would allow for continuous four-dot grids on 280mm x 356mm (11“ x 14”) film. . The production method is as follows.