论文部分内容阅读
We are currently investigating the spatial resolution of highly pixelated Cadmium Zinc Telluride(CZT)detector for imaging application. A 20 mm×20 mm×5 mm CZT substrate was fabricated with 600 μm pitch pixels(500 μm anode pixels with 100 μm gap) and coplanar cathode. Charge sharing between two pixels was studied using collimated a 122 keV gamma ray source. Experiments show a resolution of 125 μm FWHM for double-pixel charge sharing events when the 600 μm pixelated and 5 mm thick CZT detector biased at-1000V. In addition, we analyzed the energy response of the 600 μm pitch pixelated CZT detector.
A 20 mm × 20 mm × 5 mm CZT was fabricated with 600 μm pitch pixels (500 μm anodes pixels with 100 μm gap) and Coplanar cathode. Charge sharing between two pixels was studied using collimated a 122 keV gamma ray source. Experiments show a resolution of 125 μm FWHM for double-pixel charge sharing events when the 600 μm pixelated and 5 mm thick CZT detector biased at-1000V. In addition, we analyzed the energy response of the 600 μm pitch pixelated CZT detector.