论文部分内容阅读
血细胞分析是医学实验室最常采用的检验项目之一,不仅对疾病的诊断、鉴别诊断、疗效观察和预后分析有重要的临床意义,而且对健康状态的评估也有一定价值。美思康MC-6200全自动血细胞分析仪是深圳市美思康电子有限公司生产的,该仪器因价格便宜,操作方便,在县级以下医院应用较广泛。但是由于县乡级医院位置偏僻,交通不便,如果出现故障,工程师不能及时赶到现场,本文通过详细介绍美思康MC-6200全自动血细胞分析仪在使用过程中常见的堵孔故障、气泡故障、HGB报警、真空度报警、本底不正常、WBC过高等故障,分析其工作原理及原因,指出故障解决思路和方法,并给出一些仪器日常保养和维护的建议,能降低故障发生率。检验师可根据提示自行解决一些常见故障节省时间,提高工作效率;同时,也为维修工程师提供一些参考案例。
Blood cell analysis is one of the most commonly used test items in medical laboratories. It not only has important clinical significance for disease diagnosis, differential diagnosis, curative effect observation and prognosis analysis, but also has certain value for the assessment of health status. Meixikang MC-6200 automatic hematology analyzer is produced by Shenzhen Meisikang Electronics Co., Ltd. The instrument is widely used in hospitals below the county level because of its cheap price and convenient operation. However, due to the remote location of the county-level hospitals, traffic inconvenience, if failure occurs, engineers can not be rushed to the scene in time, this article describes in detail the Meisikang MC-6200 automatic hematology analyzer in the use of common plugging failure, bubble failure , HGB alarm, vacuum alarm, background abnormal, WBC too high fault, analyze its working principle and reason, point out the solution ideas and methods of fault, and give some advice of routine maintenance and instrument, can reduce the failure rate. Inspectors can solve some common problems based on tips to save time and improve work efficiency; the same time, maintenance engineers also provide some reference cases.