【摘 要】
:
1、故障分析及排除方法做心电图时,发现信号衰减与失真较严重,做出来的心电图与正常相差甚远,将导联选择开关置“0”,启动走纸开关,按“1mV”定标,走纸上方波正常,增益可调,说明主机
论文部分内容阅读
1、故障分析及排除方法做心电图时,发现信号衰减与失真较严重,做出来的心电图与正常相差甚远,将导联选择开关置“0”,启动走纸开关,按“1mV”定标,走纸上方波正常,增益可调,说明主机正常,周围环境无干扰源。此故障可能来自导联线。用DT890A万用表“200”档测导联各连线的首
1, failure analysis and exclusion methods When doing ECG, found that the signal attenuation and distortion is more serious, the ECG is far from the normal difference, the lead selection switch set “0”, start the paper switch, press “1mV ”Scaling, the normal wave above the paper, adjustable gain, indicating that the host is normal, no interference from the surrounding environment. This fault may come from lead wires. Use DT890A Multimeter“200” to measure the lead of each lead