人体血液或尿中K~+、Na~+的含量直接反映了某些器官的功能或病变。例如失常的K~+量,能引起器官如心脏、神经系统和肌肉的功能失调。早先在医院常规分析血液中钠、钾含量常采用原子吸收光谱法和火焰光度法等有焰分析,国外八十年代初出现微机化的钾钠分析仪。本文介绍的S671A型血液钾钠分析仪是采用微机控制和数据处理的半自动测量仪器,能测量人体的全血、血清和血浆中钾钠离子的含量,不需要对试样作预处理,具有操作简便、快速、精度高等特点,且所测结果更加接近于人体的生理状态,具有临床意义,该仪器日前通过了技术鉴定,并已由江苏电分析仪器厂投产。本仪器的测量总体结构原理见图1。在微机控制下对同步电机(蠕动泵)进行驱动控制,采样并输入Na~+、K~+电极测得电位值,进行浮点数字运算,并将结果进行LED数字显示。
Human blood or urine K ~ +, Na ~ + content directly reflects the function of some organs or lesions. Such as abnormal K ~ + amount, can cause organs such as heart, nervous system and muscle dysfunction. Earlier in the hospital routine analysis of blood sodium, potassium content often used flame atomic absorption spectrometry and flame analysis method, the early eighties abroad appeared potassium sodium analyzer. This article describes the S671A blood potassium sodium analyzer is a computer-controlled and data processing semi-automatic measuring instrument that can measure the body’s blood, serum and plasma potassium-sodium content, without the need for pretreatment of the sample, with the operation Simple, rapid, high accuracy, and the measured results closer to the physiological state of the human body, has clinical significance, the instrument has passed the technical appraisal, and has been commissioned by Jiangsu Electric Instrument Factory. The overall structure of the instrument measuring principle shown in Figure 1. Under the control of the microcomputer, the synchronous motor (peristaltic pump) is driven and controlled, and the potentials measured by Na ~ + and K ~ + electrodes are sampled and input, floating-point digital operation is performed, and the result is digitally displayed.