论文部分内容阅读
医用热力灭菌设备作为医疗卫生、食品生产、检验检疫等行业的重要设备,其技术参数中的温度参数影响灭菌效果。根据多年高压灭菌器校准经验,参照JJF 1308-2011《医用热力灭菌设备温度计校准规范》对工作中常见的医用灭菌热力设备温度偏差的测量结果进行不确定度评定。
Medical thermodynamic sterilization equipment is an important equipment in industries such as medical and health care, food production, inspection and quarantine, etc. The temperature parameters in the technical parameters affect the sterilization effect. According to many years of experience in autoclave calibration, reference to JJF 1308-2011 “Medical thermodynamic sterilization equipment thermometer calibration specification” on the work of medical sterilization thermodynamic equipment temperature deviation measurement uncertainty assessment.