我院是一所有160张病床的综合性县级医院。放射科现在使用的X线机,其中一台XG200mAX线机是1980年上海生产的产品、自安装投入使用以来,其性能基本良好。但由于工作量大(平均每天上消化道造影5人次之多),在使用过程中也由于电源电压不够稳定,工作人员操作时不够细心,以及机器零部件本身的自然老化,致使某个部位发生故障也是必然的。一、故障现象当机器接通电源启动,使用50mA及小焦点5mA透视档时,竟出现过载指示信号,当过载故障消除后,下球管点片试验不曝光。
Our hospital is a comprehensive county-level hospital with 160 beds. X-ray machines currently used by radiology department, of which one XG200mAX wire machine was manufactured in Shanghai in 1980, has basically performed well since it was installed and put into use. However, due to heavy workload (an average of more than 5 times daily upper gastrointestinal radiography), but also in the use of the process because of the power supply voltage is not stable enough, the staff not careful operation, and the natural aging of the machine parts, resulting in a site Fault is inevitable. First, the fault phenomenon When the machine is powered on, the use of 50mA and small focus 5mA perspective file, actually appeared overload indication signal, when the overload fault is eliminated, the next ball point film test is not exposed.