论文部分内容阅读
例1故障现象:有圆消隐,透视时屏幕上无图像。故障分析:X线机有圆消隐信号,说明中心控制器工作正常,中心控制器到监视器的电缆连接良好。故障可能产生于:豆.X光源;2.影像增强器;3.电视摄像机。故障检查:首先检查X光源和影像增强器。卸下摄像机,从增强器的输出窗口
Example 1 Symptom: There is a circular blanking and there is no image on the screen during perspective. Failure Analysis: X-ray machine has a circular blanking signal, indicating that the central controller is working properly, and the cable connection from the center controller to the monitor is good. Failure may result from: beans. X light source; 2. Image Intensifier; 3. TV camera. Troubleshooting: First check the X-ray source and image intensifier. Remove the camera from the booster output window