摘要目的评估定位像对胸部总辐射剂量的影响,包括定位像剂量和随后应用管电流调制技术(TCM)进行胸部CT检查的剂量。材料与方法采用带和不带有不同尺寸乳腺附件的192层CT扫描模拟人体模型,进行前后(AP)或后前位(PA)定位像检查和应用TCM的胸部CT检查。基于获得的CT数据,通过蒙特卡罗模拟得到剂量分布。采用蒙特卡罗模拟的定位像扫描时,球管位置固定在0°和180°;胸部CT采用TCM技术的螺旋扫描。通过采用固定起始角(0°、90°和180°)的TCM曲线和蒙特卡罗模拟来研究球管起始角度对
Abstract Objective To evaluate the impact of positioning images on the total radiation dose of the thorax, including the imaging dose and subsequent chest CT using tube current modulation (TCM). MATERIALS AND METHODS A 192-slice CT-scan model of the human body with and without mammillary accessories of different sizes was used to perform anterior and posterior (AP) or posterior (PA) imaging and chest CT examination using TCM. Based on the CT data obtained, the dose distribution was obtained by Monte Carlo simulation. With Monte Carlo simulation of the positioning image, the position of the tube was fixed at 0 ° and 180 °, and the chest CT was scanned using a TCM technique. The initial angle of the bulb was studied by using TCM curves and Monte Carlo simulations with fixed starting angles (0 °, 90 ° and 180 °)