髋关节模拟试验机在多方性复合运动方式、恒定载荷和血清润滑条件模拟了炭/炭复合材料人工髋关节在临床应用中的摩擦学行为.采用扫描电镜、能谱和激光粒度分析仪分析了其磨损表面状态和磨损颗粒特性.结果表明,医用炭/炭复合材料人工髋关节在不同碳纤维方向呈现不同的磨损特性,其磨损颗粒具有四种形貌并且尺寸分布在亚微米至数十微米之间,其磨损机理主要是微切削作用和磨粒磨损.
The hip joint simulator was used to simulate the tribological behavior of the C / C composite hip joint in clinical application under the method of multi-complex movement, constant load and serum lubrication.Using scanning electron microscopy, energy dispersive spectroscopy and laser particle size analyzer The wear surface state and the wear particle characteristics of the medical hip joint.The results show that the medical hip joint of the medical carbon / carbon composite material has different wear characteristics in different carbon fiber directions, the wear particles have four morphologies and the size distribution is in the submicron to tens of micrometers Between the wear mechanism is mainly micro-cutting and abrasive wear.