《医院消毒供应中心管理规范》之清洗消毒及灭菌效果监测标准(WS310.3-2009)中指出:预真空压力蒸汽灭菌器(包括脉动真空)应在每天开始灭菌前进行B-D测试,B-D测试合格后,灭菌器方可使用。常规B-D测试是在灭菌柜选预热后空载状态下只将B-D试验包置于灭菌柜内灭菌车的前底层,靠近柜门与排气口底前方〔1〕处再运行BD程序,B-D程序结束后B-D测试合格,证实灭
The Standard for Sterilization and Sterilization Monitoring of Hospital Disinfection Supply Center (WS310.3-2009) states that pre-vacuum steam sterilizers (including pulsating vacuum) should be tested for BD prior to the start of sterilization each day, BD test after the sterilizer before use. The conventional BD test is to put the BD test pack in the sterilizing cabinet only in the front bottom of the sterilizing vehicle under the no-load state after the pre-heating of the sterilizing cabinet is selected, and then run the BD near the front of the cabinet door and the exhaust outlet [1] Program, BD test after the end of the BD program, confirmed off