进口生化分析仪和电生理仪器中,常使用轻触控制面板。这种面板操作方便,便于清洁。但操作人员使用不当,使其常受重力及尖状物触压,或仪器长期使用后,面板的接触点也要衰老损坏,造成仪器功能失控。该面板厚度约1mm,可分解成如图所示的5层。第①层印有各键功能的文字;第②层是弹性隔离层;第③层和第⑤层是上下对应的接点组及输出连线。在这两层之间,隔有一层弹性薄膜(第④层)。在正常情况下,按键松手后,第②层薄膜以其弹性将第①层复原,而第④层的弹性则使第③和第⑤层触点分离。
Imported biochemical analyzer and electrophysiological instruments, often use the touch control panel. This panel is easy to operate and easy to clean. However, improper use of operators to make it often under the pressure of gravity and spikes, or long-term use of the instrument, the contact point of the panel also aging damage, causing the instrument out of control. The panel thickness of about 1mm, can be decomposed into five layers as shown. The first layer is printed with the text of each key function; the second layer is a flexible isolation layer; the third layer and the fifth layer is up and down the corresponding contact group and the output connection. Between the two layers, separated by a layer of flexible film (the fourth floor). Under normal circumstances, the button to let go, the second layer of the film with its flexibility to recover the first layer, while the fourth layer of the elasticity of the third and fifth layer contact separation.