论文部分内容阅读
丹麦 ABL_3血气分析仪打印机常见故障有:打印机驱动马达不转,打不出数据打出的数据模糊或缺少笔划。由于该打印机体积小,结构紧凑,且多无维修资料可供参考,因而维修极不方便。下面介绍故障维修一例。故障现象是,马达运转正常,并能按指令打出数据,但字符模糊,有的字符缺少笔划。据现象分析可知,马达驱动电源及打印机运行指令系统均不会有何问题,故障很可能出在打印头、打印带及其它机械部件上。先换打印带试验,结果故障与打印带无关。再取
Danish ABL_3 blood gas analyzer printer common faults are: the printer driver motor does not turn, hit the data can not play the data fuzzy or lack of strokes. Due to the small size of the printer, compact structure, and more maintenance-free information for reference, so maintenance is extremely inconvenient. Here is an example of fault repair. Fault phenomenon is that the motor is running properly, and can press the command data, but the characters are fuzzy, and some characters are missing strokes. According to the phenomenon analysis shows that the motor drive power and printer operation command system will not have any problems, the failure is likely to print head, print tape and other mechanical parts. First test with the print tape, the results of the fault has nothing to do with the print. Then take