动脉血液中的氧分压对于评价许多急性和住院病人的病情是一项重要的测试指标。现行情况下该指标的测试都是进行动脉血液抽样,而后在临床实验室进行手工或自动的分析(使用Clark电极)。不尽人意的是,这种单一的采样分析方法耗费较高,且不能在动态变化的临床环境下提供实时信息。本文研究了一种新型的可应用于临床的光纤式氧传感器。该传感器的指示器在短暂的UV刺激后可变成易于被吸收,尔后又返回到透明状态。指示器返回到透明状态的速率正比于局部氧浓度。指示器的吸收性能可用一红色的发光二极管
The partial pressure of oxygen in arterial blood is an important test indicator for evaluating the condition of many acute and hospitalized patients. The current test of this indicator is to take arterial blood samples and then perform manual or automatic analysis in clinical laboratories (using Clark electrodes). Unfortunately, this single sample analysis method is expensive and does not provide real-time information in a dynamically changing clinical setting. This paper studies a new type of fiber optic oxygen sensor that can be applied to clinic. The sensor’s indicator can become readily absorbed after a brief UV stimulus and then return to a clear state. The rate at which the indicator returns to the transparent state is proportional to the local oxygen concentration. A red light-emitting diode can be used for the absorption performance of the indicator