目前进口气相色谱仪广泛使用电子流量、电子压力控制技术,可精密控制仪器各种载气流速,提高仪器的灵敏度和测试的重现性,使仪器的智能化更高,设计和维修也更为复杂,在AFCAPC故障时,厂家通常的做法是整体更换该模块及捕集阱,因该模块结构精密设计复杂,整套更换的代价很高,往往在2-3万元。一直以来都被很多用户和仪器工程师误以为是不可维修的,主要是因为其相关资料太少,加之接触的少,因此对该模块的组成和工作原理也是模糊不清。笔者已多次处理过各品牌气相色谱仪的自动流量控制故障,针对这种认识上的误解,现特通过岛津GC2010 PLUS AFC维修实例,以便让大家对电子流量控制的工作原理有一个直观的认识,增强我们勇于实践的信心,打破其不可维修的传言。
At present imports of gas chromatography widely used electronic flow, electronic pressure control technology, precision control of various carrier gas flow rate, improve instrument sensitivity and test reproducibility, the instrument more intelligent, more design and maintenance Complex, the AFC APC failure, manufacturers usually replace the module and the trap as a whole, due to the complex structure of the module sophisticated design, replacement of the entire package is very expensive, often in 2-3 million. It has been mistaken by many users and instrumentation engineers as being unmaintainable, mainly because of the paucity of relevant information and the lack of contact, and thus the composition and working principle of the module are also obscure. The author has repeatedly dealt with the automatic flow control failure of each brand of gas chromatograph, for this understanding of the misunderstanding, are now available through the Shimadzu GC2010 PLUS AFC maintenance example, so that everyone on the electronic flow control works have an intuitive Understand and enhance our confidence to practice, to break the rumor that they can not be repaired.